Product Description
ASTM E1491 Dust Cloud Minimum Ignition Temperature Tester Dust Cloud MIT-cloud Test
1.Main functions
The instrument is used to test the lowest ignition temperature of dust cloud and evaluate the potential explosion risk of spontaneous combustion of dust cloud. This device is in accordance with national standardsGB/T16429 as the basis, the use of modern advanced technology, mechanical, electronic and computer technology in one, can automatically test the most dust cloud Low ignition temperature. The instrument uses intelligent temperature control system, reasonable structure, stable performance, simple operation, is an ideal analysis and testing equipment.
ASTM E1491-06 Standard Test Method for Minimum Autoignition Temperature of Dust Clouds
EN 50281-2-1:1999 Electrical apparatus for use in the presence of combustible dust. Test methods Methods of determining minimum ignition temperatures.
IEC 61241-2-1:1994 Electrical apparatus for use in the presence of combustible dust - Part 2: Test methods - Section 1: Methods for determining the minimum ignition temperatures of dust.
ASTM E1491 Dust Cloud Minimum Ignition Temperature Tester Dust Cloud MIT-cloud Test
2.Scope of application
1. Determination of the lowest ignition temperature of dust cloud.
2. Evaluate the potential explosive risk of spontaneous combustion of dust clouds.
3. Combustible dust that relies on oxygen in the air to maintain its chemical reaction.
ASTM E1491 Dust Cloud Minimum Ignition Temperature Tester Dust Cloud MIT-cloud Test
3.Technical indicators
1. Test system
1.1 Device structure: standard quartz test tube, dust cover and gas storage tank
1.2 Detection mode: photoelectric automatic detection of fire
1.3 Dust spraying pressure: 0 ~ 0.1MPa (gauge pressure) Built-in air pump
2. Integrated control and data acquisition and processing system:
2.1 Microcomputer automatic control:
After input control parameters through the keyboard and display of the machine and start the test, the instrument will automatically test according to the set value
To determine whether spontaneous combustion occurs, the whole experiment is completed automatically.
2.2 Computer Remote wireless/wired monitoring (optional)
After input control parameters through the monitoring software operation interface and start the test, the instrument automatically completes the whole experiment according to the set value.
Parameters are automatically collected and stored in data files.
2.2.1 Computer remote wireless monitoring, data wireless transmission and storage, man-machine separation, to ensure the personal safety of test personnel.
2.2.2 Computer remote wired monitoring.
2.3 Video recording: Motion camera can be configured to automatically record experiment video (optional).
3, temperature control system:
3.1 Temperature control range: room temperature ~ 1000ºC Temperature control accuracy: ±2ºC
3.2 Temperature measuring part: middle of heating furnace
3.3 Temperature control part: the whole heating furnace
3.4 Temperature control sensor: Type K thermocouple
Temperature detection range: 0 ~ 1000ºC Resolution: 0.1ºC Accuracy: 0.25%FS
4. Safety protection:
4.1 Computer remote wireless monitoring, data wireless transmission and storage, man-machine separation, to ensure the personal safety of test personnel (optional).
4.2 Configuring the emergency stop switch, press the emergency stop switch quickly in an emergency to cut off the power supply of the device to ensure the safety of the device and experimental personnel.
5, display: color capacitive touch screen
6. Environmental monitoring
Atmospheric pressure measurement range |
80 ~ 110.0kPa Resolution: 0.1kPa |
Ambient temperature measurement range |
-40 ~ 80ºC Resolution: 0.1ºC |
Ambient humidity measurement range |
0 ~ 100%RH Resolution: 0.1%RH |
Power source |
220V±10% AC 50Hz±2Hz Power: 2500W |
Working environment |
Ambient temperature: 10 ~ 35ºC Ambient humidity: 20% ~ 85%RH |
equipment size |
78cm*58cm*75cm |
equipment weight |
50KG |
installation requirements |
fume hood and industrial vacuum cleaner (user provided) |
ASTM E1491 Dust Cloud Minimum Ignition Temperature Tester Dust Cloud MIT-cloud Test
4.Standard fittings
1 host 1 set
2 Test furnace 1
3 Pressure sensor (built-in) 1
4 K type thermocouple 1
5 One combustion detector
6 One viewer (built-in)
7 Glass adapters 2
8 Connect five silicone hoses
9 One airway
10 Dust bin 1 set
11 Demo dust 1 bag
12 One copy of instruction manual
13 Packing list 1 copy
14 Certificate 1
15 Five fuses
16 One power cable
ASTM E1491 Dust Cloud Minimum Ignition Temperature Tester Dust Cloud MIT-cloud Test
5.Select accessories
1. Laptop (including monitoring software) 1 set
2. 1 set of HY433-WL wireless data transmission module
3. 1 data cable (3 meters)
4. Motion camera (including control module)
Detailed Photos




A 2850K 2×60W Incandescent Sun-settting Light Yellow Li