Customization: | Available |
---|---|
After-sales Service: | Online |
Warranty: | 1 Year |
Suppliers with verified business licenses
LC-6050 High Performance Liquid Chromatograph Liquid Chromatography Equipment
Description:
The LC-6050 high-performance liquid chromatograph extensively uses high-quality components, enabling it to handle long-term high-load detection tasks and significantly enhancing its durability. Combined with a high-precision automatic sample introduction system, it achieves rapid chromatographic analysis.
LC-6050 is a liquid chromatography system with outstanding scalability and compatibility. It can meet the diverse needs ranging from routine analysis to highly efficient and rapid analysis, providing users with continuous and stable analytical results. By establishing network communication, it enhances the analysis efficiency and offers more possibilities for the rational layout of the laboratory.
Applications
Liquid chromatography is a biochemical analysis instrument widely used in various industries. It separates mixtures by utilizing the difference in the distribution ratio of the mixture between a liquid and two immiscible liquids, and then conducts analysis and identification.
Life Sciences: Techniques for the separation, analysis and purification of biological macromolecules.
Medicine: Drug analysis, synthesis, quality control and metabolic research.
Food: Detection of food additives, pesticide residues and nutritional components.
Environmental monitoring: Analysis of pollutants in water bodies, atmosphere and soil.
Chemical Engineering: Quality control of raw materials, products and fine chemical products.
Education and Research: As a common analytical tool for teaching and research.
LC-6050 High Performance Liquid Chromatograph Liquid Chromatography Equipment
Module | Parameter | Specification |
Pump | Degasser | 6-channel online, high-performance tubing |
Purge Function | One-key bubble purge | |
Solvent Channels | Binary or Quaternary (optional) | |
Pressure Range | 0-660 bar (9,600 psi) with auto-alarm/shutdown | |
Mixer Volume | 400 μL (1 mL optional) | |
Mixer Pressure Rating | 1,000 bar (14,500 psi) | |
Autosampler | Cycle Time | ≤8 s (full injection) |
Tray Capacity | 108 positions (1.5 mL vials) | |
Injection Volume | 0.0-100.0 μL (customizable) | |
Max Pressure | 450 bar (6,500 psi) | |
Carryover | ≤0.0005% (caffeine @273nm) | |
Precision (RSD) | ≤0.15% | |
Needle Wash | Online degassing, programmable speed | |
Column Oven | Heating Technology | Direct ceramic heating |
Column Capacity | 4×250mm columns | |
Temperature Range | Ambient +3°C to 85°C | |
Solvent Tray | 4×1000mL + 2×500mL bottles | |
UV-Vis Detector | Wavelength Range | 185-700nm or 185-900nm |
Light Source | D2 & W lamps (side-access) | |
Bandwidth | 8nm (4nm optional) | |
Sensitivity | 2×10-5g/mL naphthalene (with column) | |
Noise/Drift | ≤1×10-5AU / ≤3×10-5AU | |
Flow Cell | 9.5μL (3.9/6.4μL opt.), 90bar, 10mm path (7mm opt.) | |
Data Rate | 80 Hz | |
Modes | Static scan & dynamic programming | |
DAD Detector | Wavelength Range | 200-800nm |
Diode Array | 512 elements | |
Flow Cell | 8μL (1/16μL optional) | |
Noise/Drift | ≤1×10-5AU / ≤3×10-5AU/h | |
Analog Outputs | 4 channels | |
Data Rate | 100 Hz | |
RID Detector | Principle | Refractive index (1.00-1.75 RIU) |
Range/Linearity | 0.2-512 μRIU / 600 μRIU | |
Noise/Drift | 2.5 nRIU (τ=1.5s) / ≤200 nRIU/h (water, purge off) | |
Response Times | 0.1-6 s (7 steps) | |
Auto-zero | Optical/electronic |