Product name | Two van hot and cold impact testing machine Use of low temperature and high temperature hot and cold storage tank, in accordance with the cylinder, the hanging basket in turn to high temperature and low temperature box, so as to achieve rapid temperature shock effect, balance (BTC) + special temperature control system design of supply air circulation system, to control SSR P.I.D. way, make the system of heat equal to the amount of heat loss, thus the use of long-term stability. |
|||||
Product model | TSD - 11 - a - 2 p (air-cooled type) Note: air-cooled type need not install cooling towers, water cooled type to install, depends on customer's requirement. |
|||||
Limit sample | The test equipment prohibited: (1)Which is inflammable, explosive, easy to volatile substances sample test or store (2) the corrosive substance sample test or store (3) biological sample test or store (4) strong electromagnetic emission source sample test or store |
|||||
Volume, size and weight | ||||||
Nominal content product | 52L | |||||
Shopping basket size | W50*H32*D32cm | |||||
Installation dimensions | W132*H187*D155CM | |||||
weight | About 400kg | |||||
Performance | ||||||
Test environment conditions | Environment temperature is 25 ºC, relative humidity of 85% or less, test under the condition of no sample in the box GB/T 5170.2 1996 temperature testing equipment. |
|||||
High greenhouse | +60ºC→+150ºC | |||||
Time | +60ºC→+150ºC ≤25min | |||||
Temperature range | -55ºC~-10ºC | |||||
Time | +20ºC → -55ºC≤60min | |||||
Test (style) | Test chamber upside for high temperature, lower for the low temperature zone, specimen is placed in the basket, basket turns in the high temperature box and low temperature in experiment | |||||
Shock temperature range | (+60~+100)ºC/(-40~-10)ºC | |||||
Temperature fluctuation degree | ±0.5ºC | |||||
The temperature deviation | ±2.0ºC | |||||
The temperature recovery time | ≤5min | |||||
Temperature conversion time | ≤10sec | |||||
Restore conditions | High temperature exposure: + 150 ºC / min - > low temperature exposure: 30 to 55 ºC / 30 min Sample maximum weight: 2.5 KG |
|||||
Meet the test standard | GB/T 2423.1 2008 test: A test method at low temperature GB/T 2423.2 2008 test B: high temperature experiment method GB/T2423.22-2002 test N: temperature change test method test Na GJB 150.5 A - 2009 mili tary equipment laboratory test method for part 5: the temperature impact test GJB 360 a - 1996 temperature impact test GJB367.2-2001 mili tary communication equipment general specification temperature impact test IEC 68-2-14 test method for N: temperature changes |
|||||
Structural features | ||||||
Thermal insulation enclosure structure | (1) the outer wall material: advanced electrolytic plate surface treatment of the lacquer that bake. (2) the lining material: stainless steel plate SUS 304. (3) low greenhouse body heat preservation material, high temperature resistant rigid polyurethane foam. (4) high greenhouse body heat preservation material: glass fiber high temperature resistant white cotton. (5). The door is thermal insulation material: glass fiber |
|||||
Air conditioning channel | (1) loop motor: 2 1/2 HP (2) heater, evaporator, pressure balance mouth, adjustable wind deflector, temperature sensor. |
|||||
Test cases of standard configuration |
(1) plane embedded handle, (2) the door hinge: SUS # 304 import hinge (3) SUS # 304 stainless steel grille type 2 slices steel shelf, single biggest bearing 10 kg. (4) stainless steel SUS # 304 can adjust the spacing between 40 mm shelf track 2 groups (5) lead hole: phi 50 mm one |
|||||
Door |
(1) single door, to open outward, hinges on the left, put his hand on the right (in the face of cabinet positive). (2) the window/door frame prevent condensation heating device. |
|||||
The control panel | 7 inch controller display set and overtemperature protection device, USB and RS232 communication interface | |||||
Machine room | Refrigeration units, compressor water pans, exhaust pressure device | |||||
Power distribution control cabinet | Distribution board, cooling fan, the total power switch device | |||||
The heater | (1) the fin type radiator pipe shaped stainless steel electric heater (2) the heating control mode: SSR (solid state relay) non-contact cycle such as pulse width (3) the low temperature district heating power: 3.0 KW (4) high temperature heating power: 6.0 KW |
|||||
The power cord and drain holes | located on the back of housing | |||||
Refrigeration system | ||||||
Cold and hot device | Overlapping type in Europe and the high efficiency power-saving type cryogenic refrigeration system (air-cooled heat). | |||||
Hot and cold switching device | Adopt high efficiency coal SWEP cold cold heat exchange design, system of Sweden (316) sus than traditional spiral in high efficiency. | |||||
Heat load regulation | (1) mining computer automatic cold coal flow, effectively take away heat load and under test DE/day Synchronization operation. (2) than the traditional design, high control stability and reproducibility and to save electricity, low humidity of the super efficiency. |
|||||
Refrigeration compressor | French taikang efficient full close compressor at low temperature (6 HP * 2 sets) | |||||
The efficient components | (1) condenser, evaporator mining AC&R compound vortex type aluminum fins. (2) the enterprisesin throttling device adopts imported refrigeration fittings. |
|||||
Expansion system | The capacity control of the refrigeration system (capillary). | |||||
Evaporative condenser | Stainless steel welded plate heat exchanger. | |||||
Machine control mode | (1) the controller of the control system according to the test conditions automatically adjust the operating conditions of refrigerating machine. (2) the evaporating pressure regulator. (3) compressor gas cooling loop back (4) energy regulatory circuits (5) condensation pressure regulator (6) machine system design such as shock absorber tube vibration and noise reduction measures |
|||||
Refrigerating fluid | Honeywell R404a (ozone loss index of 0)/R23 environmental protection refrigerants | |||||
Electrical control system | ||||||
The controller model | South Korea three yuan import controller | |||||
Operation mode | SPC | |||||
Set way | Chinese and English menu, touch screen input | |||||
Set the range | (1) high greenhouse temperature limit: + 190 ºC (2) low greenhouse precooling temperature limit: - 75 ºC (3) laboratory (sample) : the temperature shock cap + 10 ºC, the temperature impact floor - 10 ºC |
|||||
e) The controller specification | Accuracy: the plus or minus 0.1 ºC temperature + 1 digit Resolution: plus or minus 0.01 ºC temperature Temperature slope: 0.1 ~ 9.9 can be set. Standby has upper and lower limits and alarm functions. The temperature signal into force T armoured thermocouple sensor. 6 set P.I.D control parameters setting, P.I.D automatic calculus. |
|||||
f) Image display | In picture dialogue without keystroke, direct touch screen option. Temperature setting (SV) and the actual value (PV) display directly. It can display the current enforcement procedure number, period of time, the remaining time and cycle times. Running total time function. |
|||||
g) Capacity and control function | Can use the program group: biggest 100 PATTEN. Repeatable execution procedures: each program cycle of up to 9999 times. Program type and the production of mining dialogue, with functions of editing, remove and insert. SEGMENTS hour59min time setting 0 ~ 540. After power with procedural memory, complex start automatically and the function in the executable program. / RS - 485 communication interface with the function of USB interface. When the program execution curve can be real-time display graphics. With functions of automatic adjust refrigerating capacity. With functions of reservation startup and shutdown. Buttons and screen LOCK (LOCK) function. Troubleshooting tips showed. The screen can be adjusted for backlight. Screen protection for timing, closed set the TIMER or manually. |
|||||
Safety device | ||||||
Refrigeration system | Compressor flow compressors overheating Low pressure compressor ehv/compressor The cooling fan overheating/owe phase protection |
|||||
Test chamber | Temperature too high or too low to protect/adjustable overtemperature protection device Air conditioning channel overtemperature protection/cycle motor overheating protection |
|||||
other | Total phase sequence and open phase protection/power leakage protection/heating tube without dry burned protection Overload and overheating protection/protect/compressed air power switch/motor temperature switch/lightning protection device |
|||||
The choose and buy accessories (function) of the following attachments (function) quoted separately, need special specification in the contract or the technical agreement | ||||||
Pin hole | Pin holes have a diameter of 50 mm phi, 80 mm, 100 mm, phi phi phi 120 mm specifications, its position and quantity structure allows the enclosure and does not affect the performance of the conditions can be customized according to the requirements of customers | |||||
Heavy-duty rack | Have weight 50 kg/layer, 80 kg/layer, 100 kg/layer (the rack in the oven aggregate carrying no more than 200 kg) | |||||
recorder | The choose and buy: the paperless recorder.(with USB interface) | |||||
Centralized monitoring software | For remote centralized monitoring, along with the software (centralized monitoring software) with the RS - 485 / RS - 232 converter 1 Software using the environment: IBM PC compatibles, P above CPU, more than 128 MB of memory, simplified Chinese Windows 128 or simplified Chinese Windows xp operating system |
|||||
l) The remote control | LAN monitoring system (options) | |||||
Transport | Test chamber for integral, and the overall transportation | |||||
size weight |
Biggest transport size (including packaging) : see "dimension" Biggest shipping weight (including packaging) : see "weight" |
|||||
Conditions of use | Ensure that the following conditions by the user | |||||
The installation site | The ground level off, the well ventilated Equipment around without strong vibration Without strong electromagnetic field surrounding equipment Equipment around no inflammable, explosive, corrosive substances and dust With proper use and maintenance space around the equipment |
|||||
Environmental conditions | Temperature: 5 ºC ~ 35 ºC (2) the relative humidity: 85% or less (3) pressure: 86 kpa to 106 kpa t | |||||
p)The power supply Power Maximum current The power switch |
AC380V three phase four wire + ground protection Voltage allowed range: AC (380-38) V Frequency allowed range: (50 + / - 0.5 Hz Protect the earth grounding resistance less than 4 Ω Requires the user to the installation site of air or power switch device configuration corresponding capacity, and the switch must be independent for the use of this equipment About 12.0 KW 18 a 30 a |
|||||
air compression | Because the valve action on compressed air, driven air compressor pressure shall not be less than 5 kg/CM2. | |||||
Requirements for storage environment | When the environment temperature is lower than 0 ºC, equipment should be kept clean water discharge, so as not to pipe water freeze up bad pipe (except the air-filled type machine) | |||||
Technical training | Free for your company training of operation and maintenance personnel (location: at the scene of the expensive department) | |||||
Coordination &liaision instead | 10 days (spot, to inform the factory before the acceptance of the contract become effective) | |||||
After-sales service | Since the date of transfer of equipment, one year free warranty, equipment is not regular visit, life-long service and provide spare parts for a long time. | |||||