Customization: | Available |
---|---|
After-sales Service: | Online |
Warranty: | 1 Year |
Suppliers with verified business licenses
Volume and dimension | ||||||
Effective volume | 80L | |||||
Inner box size | W500×H400×D400 mm | |||||
Outer box dimension |
W1400×H1950×D1600 mm | |||||
Performance indicators (ambient temperature +5ºC~30ºC under no-load conditions) | ||||||
Temperature range | -80ºC~+220ºC | |||||
Impact temperature range | High temperature 60ºC ~ +200ºC Low temperature -10ºC ~ -60ºC |
|||||
Preheating temperature range | +60ºC ~ +220ºC | |||||
Heating time of high temperature tank | RT (room temperature) ~ +220ºC≤30min (room temperature +10 ~ +30ºC). | |||||
Precooling temperature range | -10ºC~-80ºC | |||||
Cooling time of cryogenic tank | RT (room temperature) ~ -80ºC≤80min (room temperature +10 ºC -- +30ºC) | |||||
Temperature fluctuation | ≤±0.5ºC | |||||
Temperature uniformity | ≤±2.0ºC | |||||
Damper transition time | ≤5S (open cylinder conversion time) | |||||
Impact recovery time | -60-- +200ºC≤5min. | |||||
Temperature exposure constant time | ≥30min | |||||
Temperature shock condition | High temperature ~ low temperature, low temperature ~ high temperature, normal temperature ~ high temperature ~ low temperature, normal temperature ~ low temperature ~ high temperature | |||||
Power source | AC3¢5W 380V 50Hz(R, STN three-phase four wire plus ground wire) (voltage fluctuation ≤±10%) | |||||
Gross power | 32KW | |||||
Performance Ambient temperature requirements | 0~+30ºC | |||||
Cooling mode | The equipment adopts air cooling mode | |||||
Energy-saving design | The refrigeration system can adjust the power according to the load size, using PID+PWM principle of VRF technology (electronic expansion valve according to the heat condition of refrigerant flow servo control), and has the compressor energy regulation technology, can achieve excellent energy saving effect | |||||
Structural material | ||||||
Inner box material | Stainless steel plate (SUS#304) Pobox, South Korea | |||||
Outer box material | Cold rolled steel plate paint | |||||
Thermal insulation material | 100mm rigid polyurethane foam + glass fiber, fire resistance, in line with environmental requirements | |||||
door | The hinged insulation door panel is one piece, the material is the same as the box. Single door with explosion-proof handle. In order to prevent condensation or frost on the edge of the door frame and door during the low temperature test, the edge of the door frame and door is provided with an electric defrost device | |||||
Sample storage rack | Stainless steel SUS304# 2 sets of adjustable space storage track, 2 pieces of grille storage rack. Each floor bears 20kg. The position of the sample holder can be adjusted as required. | |||||
seal | The door of the test chamber is sealed with silicone rubber strip, which is resistant to high and low temperature, anti-aging, and has good sealing performance | |||||
High and low temperature cylinder conversion | The use of Taiwan "ADke" professional pneumatic equipment components A. Divided into high temperature area, low temperature area, test area. The high and low temperature area is used as energy reserve, and the test area is the user's experimental area. B. The switchable valves are separated between the high and low temperature zones and the test zones. Automatically open the corresponding valve upon impact. C. There is also an independent air exchange valve, which acts to introduce ambient air during exhaust (normal temperature recovery) |
|||||
castor | The bottom of the machine adopts four sets of high-quality fixed universal movable wheels (easy to move) and flat water foot cups (adjust the balance of the body | |||||
Manufacturing characteristics | 3D full parameter modeling, intelligent and efficient; PC virtual assembly will be carried out before manufacturing to ensure that the function is perfect without design omissions; Motion simulation, stress analysis to ensure the design rationality. Channel modularization ---- ensures manufacturing consistency and stable batch performance. Joint treatment -- Japan imports high and low temperature sealant and increase stainless steel strip sealing treatment to ensure that the long-term use of the equipment as a whole does not deform, the structure does not leak. 3S manufacturing management process ---- In the factory according to fixed scientific process to ensure high efficiency, stable quality; Avoid to the customer site production, the process can not be guaranteed, resulting in unstable performance. Welding and gluing process pictures |
|||||
Air supply system | ||||||
Air supply characteristic | Air is circulated horizontally from right to left in the test chamber | |||||
Circulating motor | Taiwan Yizheng Motor | |||||
Wind wheel | Multi-wing centrifugal circulating fan with high and low temperature aluminum alloy blades | |||||
Schematic diagram of impact ventilation | ||||||
Refrigerating system | ||||||
Refrigeration mode | Cascade compressor refrigeration system | |||||
compressor | German brand "Bit zer" semi-enclosed compressor | |||||
refrigerant | Honeywell R404 refrigerant and Guangzhou Blue Sky R23 refrigerant | |||||
condenser | The high temperature and high pressure vapor discharged from the compressor is cooled into a liquid; The heat released is carried away by water or air. Can be divided into water-cooled, air-cooled | |||||
evaporator | High efficiency multi-stage aluminium alloy fin evaporator (fin thickened) | |||||
Heat exchanger | High efficiency ST plate heat exchanger | |||||
Oil separator | The Emerson oil separator separates the refrigerant and liquid sprayed out of the compressor, which is used to protect the compressor from the loss caused by the accumulation of refrigerant liquid | |||||
Filter dryer | Absorb the residual water and acidic substances in the refrigerant of the refrigeration system and filter out the solid impurity particles in the system, copper cutting, etc., protect the normal operation of the expansion valve and the capillary tube, and avoid ice blockage and water blockage | |||||
Refrigerant oil | The United States "ice bear" frozen oil, known as Uniqema frozen oil, also known as polar bear frozen oil. Ice Bear frozen oil is recognized by the world refrigeration industry as the choice of refrigeration compressor lubricating oil, has been certified by the world's major refrigeration compressor manufacturers. | |||||
High and low pressure protector | Japan "Sagomiya" brand, effectively protect the compressor to prevent irreparable damage to the compressor under high or low pressure | |||||
Electronic expansion valve | 1) Danf oss electronic expansion valve. Refrigerant flow servo control (i.e., "static balance" technology of cooling without heating and heating without cooling) 2) Electronic expansion valve, as a new type of control component, has already broken through the concept of throttling mechanism, it is an important part of the intelligent refrigeration system, but also an important means and guarantee for the real realization of refrigeration system optimization. 3) Product features: 1. Fully enclosed design, stepper motor drive 2. From full off to full on in a very short time 3. High accuracy, extremely long life 4. Flow rate change is linear, wide range of cold capacity 5. Continuous cooling capacity adjustment, no liquid hammer phenomenon in the refrigeration circuit 6. Motor and valve integrated, high reliability, ceramic valve plate and valve parts wear resistance 4) Main advantages 1. Realize the active control of the refrigeration system, instead of the fixed proportion adjustment of the original thermal expansion valve. The output can be adjusted in advance and optimally for different modes and conditions. 2. Because the closed-loop control feedback front end of the electronic expansion valve is directly controlled by temperature and pressure dual sensors with high response, the optimal evaporator liquid supply can be provided, so that the refrigeration system can achieve excellent cooling capacity output in a wider working range. 3. Energy saving: Electronic expansion valve full range of self-adjustment + active adaptation to adjust the cooling capacity output, making the system more energy saving |
|||||
Refrigeration technology | Germany "Rosenberg" electric pipe bender The low-temperature connection pipe is made of high quality oxygen-free copper pipe, and the famous German brand "Rosenberg" electric pipe bender is used for pipeline production, nitrogen filling welding, and 48-hour high pressure and leakage prevention process to ensure welding quality Pipe welding: High quality copper pipe nitrogen protection welding method, so that the copper pipe inside smooth and no oxidation, to avoid the traditional welding method caused by oxide impurities in the inner wall of the copper pipe into the refrigeration system to the compressor damage. Vibration reduction measures: the compressor bottom installation of the vibration reduction spring and anti-vibration soft cushion combination vibration reduction; Anti-vibration hoses and C-shaped elbows are added to the pipes of the refrigeration system to prevent the rupture of copper pipes caused by vibration and temperature changes, resulting in refrigerant leakage and affecting the overall performance of the system. Noise control: Low speed and high air volume condensing fan is used on the condenser to achieve low noise effect |
|||||
Defrost cycle | Refrigerated by the refrigeration system (the air in the box passes through the evaporator at a temperature lower than the dew point, and the water is precipitated) (1) Automatic setting times (2) Manually set times (3) Defrosting frequency: 20 cycles/times (at -55ºC~150ºC) |
|||||
Heating system | ||||||
Heating tube | Electric heating, high quality nickel chromium alloy electric heating wire, PID pulse width control SSR automatic adjustment of heating power. | |||||
Control mode | The output signal from the controller is realized by the SSR solid-state relay with high precision and non-contact opening | |||||
sensor | High precision "PT100" platinum temperature sensor | |||||
Control mode | ||||||
controller | "7-inch" Chinese and English true color touch screen man-machine interface (screen conversation) computer control, USB curve recording, data storage device |
|||||
Communication interface | Can be connected to computer display curve, data acquisition; Can do multiple machine synchronous control; RS-232, RS-485 and USB communication devices. |
|||||
Data recording mode | With RAM with battery protection, can save the device set value, sampling value and sampling time; Curve recording cycle can be set to 30~60sec, maximum memory time storage of 90 days Historical curve, historical data (when sampling) | |||||
Controller characteristics | Self-developed LCD touch screen direct button controller, Chinese and English 7 "graphics wide viewing Angle, high contrast with adjustable backlight function of large LCD LCD controller. 1. Controller specifications :a. Accuracy: temperature ±0.1ºC+1digit. b. Resolution: Temperature ±0.1. c. With upper and lower standby and alarm functions. d. Temperature input force signal type T. e.P.I.D control parameter setting,P.I.D automatic calculation. 2.. Screen display function :a. screen dialogue, no key input, screen directly touch options. b. The temperature setting (SV) and the actual (PV) value are displayed directly. c. Display the fault status and troubleshooting methods. d. Can display the current program number, segment, remaining time and number of cycles. e. The temperature program setting value is displayed by graphical curve, with real-time display program curve execution function. f. With a separate program editing screen, you can input temperature, time and cycle number. g. The screen can be backlit. h. Screen display protection function can be used for timing, TIMER or manual shutdown Settings. |
|||||
Protection function | ||||||
Protective alarm devic | 1. Test product overtemperature (high temperature) protection (independent, configurable on panel)2. With fusible short circuit breaker protection switch 3. Heater overtemperature overload protection switch 4. The compressor is overheated 5. Compressor high and low pressure protection 6. System overcurrent/undervoltage protection device 7. The controller determines the fault display by itself 8. Low cooling water pressure protection 9. The power supply lacks inverse phase protection, leakage and short circuit protection 10. Load short-circuit protection 11. Safety ground cable terminal 12. Cooling water supply underpressure 13. Air conditioning channel limit overtemperature 14. Fan motor overheating or overload protection |
|||||
Supplementary statement | ||||||
Installation site | Good ventilation, flat ground, in line with GB50209-2002 standard: Flatness <= 5mm/2m There is no strong vibration around the device There is no strong electromagnetic field around the device There are no inflammable, explosive, corrosive substances and dust around the equipment Leave proper space for use and maintenance around the equipment, as shown in the following figure: |
|||||
Performance Guaranteed environmental conditions | Ambient temperature: -5ºC~+28ºC; Relative humidity: ≤85%RH; Atmospheric pressure: 80kPa~106kPa | |||||
Power supply condition | Power supply 380V AC(±10%), frequency: 50 ±0.5HZ, three-phase four-wire + PGND wire, grounding resistance ≤4Ω; The air or power switch of the corresponding capacity must be configured for the device at the installation site, and the switch must be independent for the device (Recommended power switch capacity: 3P+N63A). |
|||||
Air source condition | Provided by the user (not less than 6KPA air source) | |||||
Compressor cooling system | Air cooling |
|||||
other | Opening the door of the test chamber during the test will cause the temperature in the chamber to fluctuate; During the test, if the door is opened several times or for a long time, or the test sample emits wet steam, it may cause the heat exchanger of the refrigeration system to freeze and fail to work normally. | |||||
Configuration list | ||||||
controller | Independent research and development | |||||
Ac contactor | Sch neider of France | |||||
Solid-state relay | Yangming, Taiwan | |||||
Phase sequence protection | Taiwan Shilin | |||||
Intermediate relay | Schn eider of France | |||||
transformer | Zhejiang Chint | |||||
Circuit breaker | Zhejiang Chint | |||||
Heating tube | Dongguan Wade | |||||
Electric machine | Yizheng Taiwan | |||||
Electric wire | Dongguan Mingfeng | |||||
compressor | Bit zer, Germany | |||||
refrigerant | R404 (Honeywell) | |||||
refrigerant | R23 (Guangzhou Blue Sky) | |||||
Oil separator | Emerson | |||||
Solenoid valve | Dan foss, Denmark or Casto, Italy | |||||
Expansion valve | Dan foss, Denmark | |||||
Filter dryer | Dan foss, Denmark | |||||
Suspension tube | Casto, Italy | |||||
Insulated pipe | American Paris | |||||
evaporator | Dongguan Jiaheng (customized according to use requirements) | |||||
condenser | Dongguan Jiaheng (customized according to use requirements) | |||||
High-low switch | Dan foss, Denmark | |||||
Refrigerant oil | Ice bear | |||||
Water tower | Nyaungyan | |||||
chassis | Pobin 304#, Korea |