Customization: | Available |
---|---|
After-sales Service: | Online |
Warranty: | 1 Year |
Suppliers with verified business licenses
Temperature range | Room temperature ~ 1250ºC |
Temperature resolution | 0.001ºC |
Temperature fluctuation | ±0.01ºC |
Heating/cooling rate | 0.1~100ºC/min; -0.1~-40ºC/min |
Temperature control method | PID temperature control, heating, constant temperature and cooling |
Weighing system measurement range | 0.01mg~3g, can be expanded to 50g |
Accuracy | 0.01mg |
Resolution | 0.01ug |
DSC range | ±1000mW |
DSC sensitivity | 0.0001mW |
Atmosphere control | Built-in digital flow meter, automatic control of gas size |
Software | Intelligent software can automatically record TG and DSC curves and process data. The software has an automatic adjustment function, which can automatically extend and zoom according to the spectrum display. |
Constant temperature time | Arbitrary setting, standard configuration ≤600min |
Data test frequency can be selected in real time, 2s, 5s, 10s, etc. | |
Furnace body automatic lifting and manual lifting modes, can quickly cool down; ≤15min, 1000ºC down to 50ºC | |
An external water cooling device isolates the drift of the weighing system from the heat; Temperature range -10~60ºC |